Elsevier

Clinical Biochemistry

Volume 61, November 2018, Pages 18-22
Clinical Biochemistry

Multicenter assessment of a hemoglobin A1c point-of-care device for diagnosis of diabetes mellitus

https://doi.org/10.1016/j.clinbiochem.2018.09.007Get rights and content

Abstract

Objective: A multisite investigation compared the analytical performance of a point-of-care (POC) HbA1c device with multiple commonly used HbA1c laboratory methods and an NGSP (National Glycohemoglobin Standardization Program) reference method.

Research Design and Methods: The Afinion AS100 POC device analyzed HbA1c using 618 EDTA whole blood excess patient specimens with clinically indicated HbA1c testing. Results were compared to measurements across five clinical laboratories and the NGSP reference method. Precision was evaluated over 8–10 consecutive days for low-, mid-, and high-range HbA1c specimens at all five sites.

Results: Over a wide range of HbA1c values (4.0%–15% HbA1c), 97.1% of the POC results and 94.5% of routine laboratory results fell within the target value of ±6% of the NGSP reference method results. The POC HbA1c results at 6.5% exhibited a total relative bias of −0.6% (−0.04% HbA1c) compared to the reference method while the aggregate of laboratory methods displayed a relative bias of −0.9% (−0.06% HbA1c). The total imprecision of the POC results ranged from 0.74–2.13% CV across the analytic measurement range compared to 0.81–3.23% CV for the routine laboratory methods.

Conclusions: The accuracy and precision of the Afinion POC HbA1c method was comparable to the laboratory HbA1c methods supporting the FDA's recent approval of the Afinion HbA1c Dx device for use in the diagnosis of diabetes.

Introduction

Diabetes is a significant global public health concern, with an estimated mortality rate of over 1.5 million lives per year [1]. Diabetes is a chronic condition that results from autoimmune related insulin deficiency (type 1) or insulin resistance/β-cell dysfunction (type 2), with type 2 diabetes comprising a majority of diabetes cases and new diagnoses. Type 2 diabetes results from chronic hyperglycemia that leads to insulin insensitivity at the cellular level and ultimately the inability to adequately metabolize dietary glucose [1,2].

Hemoglobin A1c (HbA1c) is formed via non-enzymatic glycation on the terminal valine of the β-globin chain. The concentration of HbA1c is highly dependent upon the average lifespan of the red blood cell and blood glucose concentrations, but is representative of an individual's average blood glucose concentration over the last 2–3 months [3,4]. Therefore, monitoring HbA1c concentrations are beneficial for assessing long-term glycemic control. Clinical guidelines from the American Diabetes Association, World Health Organization, and the International Diabetes Federation recommend that a HbA1c cutoff ≥6.5% (48 mmol/mol) can be utilized for the diagnosis of diabetes, if testing is conducted using an assay which is standardized and certified by the National Glycohemoglobin Standardization Program (NGSP) and traceable to the Diabetes Control and Complications Trial (DCCT) reference method [1,2,5]. Recently, the FDA has cleared the Abbott (formerly Alere) Afinion AS100 HbA1c Dx device as a moderately complex test for use in the diagnosis of diabetes, which is a first for a point-of-care (POC) HbA1c method.

Historically HbA1c point-of-care (POC) devices demonstrated significant bias and variability when compared to non-POC assays, which led to clinical guideline recommendations against utilization for diagnostic purposes [[6], [7], [8], [9]]. However, there have been improvements in HbA1c assays and technologies over time and further investigation into the accuracy and precision of these devices is required. POC HbA1c measurements could expedite diagnostic decisions and medical interventions provided they meet performance standards [6]. The purpose of this multicenter study was to evaluate the analytical performance of HbA1c using the Afinion AS100 POC analyzer between multiple sites, and compare the performance characteristics to other frequently used automated HbA1c assays and an NGSP reference method.

Section snippets

Instrumentation

The analytical performance of the Afinion AS100 POC analyzer (Alere Technologies AS) was compared to three clinical laboratory HbA1c platforms (Roche Tina-quant® HbA1c Gen. 3, Bio-Rad Variant II Turbo, Siemens Dimension Vista) across five clinical laboratories, and with an NGSP reference laboratory (Secondary Reference Laboratory #9, Tosoh G8 automated glycohemoglobin analyzer) (Table 1). The Afinion AS100 is a POC instrument that quantifies HbA1c in capillary or venous blood using boronate

Accuracy of POC and laboratory HbA1c values to RMP

A total of 618 patient specimens were analyzed between the POC and all laboratory methods. The Deming regression demonstrated a strong correlation (Pearson R = 0.994) between the Afinion HbA1c values and the Tosoh G8 reference method values (Fig. 1A). The mean of laboratory sites HbA1c also displayed a strong correlation (Pearson R = 0.989) with the reference method values (Fig. 1B).

Table 2 shows the total biases of the AS100 results, the aggregate mean of all local laboratory method results,

Discussion

This multicenter analytical trial demonstrates equivalent or superior performance of the Alere Afinion POC HbA1c device compared to routine clinical laboratory HbA1c assays currently employed in the diagnosis of diabetes. These results are specific to the Afinion POC HbA1c method when performed in a laboratory setting and are in accordance with previously published results [11]. The Afinion device had stronger linear correlation (Pearson R) than the aggregate of laboratory methods with less

Acknowledgements

The authors acknowledge the contribution of Alere Inc. in designing the multisite investigation, providing the Afinion AS100 point-of-care instruments to each laboratory participating in this study, and performing statistical analysis of the study results.

Duality of interest

No conflicts of interest relevant to this article were reported.

Author contributions

B.E.S. and P.M.S. conducted and oversaw the study at the University of California, San Diego Center for Advanced Laboratory Medicine under the guidance of R.L.F. A.K.S., K.S., F.S.A., M.G.S., and A.H.B.W. were responsible for conducting and overseeing the study at their respective laboratories. R.R.L. was responsible for overseeing the reference method at the DDL. B.E.S. and P.M.S. drafted the manuscript, and R.L.F., A.K.S., K.S., F.S.A., M.G.S., A.H.B.W., and R.R.L. edited and made significant

References (11)

  • World Health Organization

    Global Report on Diabetes

    (2016)
  • International Diabetes Federation Guideline Development Group

    Global Guideline for Type 2 Diabetes

    (2014)
  • I.M. Stratton

    Association of glycaemia with macrovascular and microvascular complications of type 2 diabetes (UKPDS 35): prospective observational study

    BMJ

    (2000)
  • J.W. Albers et al.

    Effect of prior intensive insulin treatment during the Diabetes Control and Complications Trial (DCCT) on peripheral neuropathy in type 1 diabetes during the Epidemiology of Diabetes Interventions and Complications (EDIC) study

    Diabetes Care

    (2010)
  • American Diabetes Association

    Standards of Medical Care in Diabetes - 2018

    Diabetes Care

    (2018)
There are more references available in the full text version of this article.

Cited by (10)

View all citing articles on Scopus
View full text