Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCT

  • Original Article
  • Published:
European Journal of Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

Purpose

The new GE Discovery STE and Discovery VCT respectively combine 16-slice and 64-slice CT with PET. The PET scanner has a new BGO detector block of 8×6 matrix (6.3×4.7×30 mm3). The aim of this study was to test the performance of the new scanner.

Methods

The PET performance evaluation was done using NEMA methodology. Owing to improved front-end electronics, the system was tested with different energy window and coincidence timing settings.

Results

Transaxial resolution FWHM for 2D(3D) mode at 1 cm offset from the centre of the field of view (R1) was 4.87 mm (5.12 mm) and at 10 cm off centre (R10) radially 5.70 mm (5.89 mm) and tangentially 5.84 mm (5.47 mm). The axial resolutions were 4.4 mm (5.18 mm) (R1) and 5.99 mm (5.86 mm) (R10). The sensitivities were 2.3 cps/kBq (8.8 cps/kBq) (R0, centre of field of view) and 2.3 cps/kBq (8.9 cps/kBq) (R10). The system scatter fraction was 21.4% in 2D at an energy of 375 keV (33.9% in 3D mode at a higher energy of 425 keV). Peak noise equivalent count rates (k=1) were 84.9 kcps at 43.9 kBq/ml (2D) and 67.6 kcps at 12.1 kBq/ml (3D). In image quality measurement the hot sphere to background contrast with 10- to 22-mm diameter spheres varied from 14% to 68%, being slightly better in 3D than in 2D mode. Cold sphere contrast was 67% in 2D and 59% in 3D mode.

Conclusion

GE’s new STE and VCT PET/CT systems have improved spatial resolution without loss in sensitivity. When compared with the LYSO crystal-based GE Discovery RX, the resolution and scatter fraction are comparable, the count rate capability is lower but the sensitivity is higher.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. DeGrado TR, Turkington TG, Williams JJ, Stearns CW, Hoffman JM, Coleman RE. Performance characteristics of a whole-body PET scanner. J Nucl Med 1994;35:1398–406.

    PubMed  CAS  Google Scholar 

  2. Brix G, Zaers J, Adam LE, Bellemann ME, Ostertag H, Trojan H, et al. Performance evaluation of a whole-body PET scanner using the NEMA protocol. J Nucl Med 1997;38:1614–23.

    PubMed  CAS  Google Scholar 

  3. Erdi YE, Nehmeh SA, Mulnix T, Humm JL, Watson CC. PET performance measurements for an LSO-based combined PET/CT scanner using National Electrical Manufacturers Association NU 2-2001 standard. J Nucl Med 2004;45:813–21.

    PubMed  CAS  Google Scholar 

  4. Brambilla M, Secco C, Dominietto M, Matheoud R, Sacchetti G, Inglese E. Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard. J Nucl Med 2005;46:2083–91.

    PubMed  CAS  Google Scholar 

  5. Kemp BJ, Kim C, Williams JJ, Ganin A, Lowe VJ. NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes. J Nucl Med 2006;47:1960–7.

    PubMed  Google Scholar 

  6. Humm JL, Rosenfeld A, Del Guerra A. From PET detectors to PET scanners. Eur J Nucl Med Mol Imaging 2003;30:1574–97.

    Article  PubMed  Google Scholar 

  7. Karp JS. Is LSO the future of PET? Against. Eur J Nucl Med Mol Imaging 2002;29:1525–8.

    Article  PubMed  Google Scholar 

  8. Surti S, Kuhn A, Werner MW, Perkins AE, Kolthammer J, Karp JS. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nuc Med 2007;48:471–80.

    Google Scholar 

  9. National Electrical Manufacturers Association. NEMA NU-2 Standards Publication NU-2-2001: performance measurements of positron emission tomography. Rosslyn, VA: National Electrical Manufacturers Association, 2001.

    Google Scholar 

  10. Daube-Witherspoon ME, Karp JS, Casey ME, DiFilippo FP, Hines H, Muehllehner G, et al. PET performance measurements using the NEMA NU 2-2001 standard. J Nucl Med 2002;43:1398–409.

    PubMed  Google Scholar 

  11. National Electrical Manufacturers Association. NEMA NU-2 Standards Publication NU-2-1994: Performance measurements of positron emission tomography. Washington DC: National Electrical Manufacturers Association, 1994.

    Google Scholar 

  12. Karp JS, Daube-Witherspoon ME, Hoffman EJ, Lewellen TK, Links JM, Wong WH, et al. Performance standards in positron emission tomography. J Nucl Med 1991;32:2342–50.

    PubMed  CAS  Google Scholar 

  13. Bettinardi V, Danna M, Savi A, Lecchi M, Castiglioni I, Gilardi MC, et al. Performance evaluation of the new whole-body PET/CT scanner: Discovery ST. Eur J Nucl Med Mol Imaging 2004;31:867–81.

    Article  PubMed  Google Scholar 

  14. Mawlawi O, Podoloff DA, Kohlmeyer S, Williams JJ, Stearns WS, Culp RF, et al. Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes. J Nucl Med 2004;45:1734–42.

    PubMed  Google Scholar 

  15. Bailey DL, Jones T, Spinks TJ. A method for measuring the absolute sensitivity of positron emission tomographic scanners. Eur J Nucl Med 1991;18:374–9.

    Article  PubMed  CAS  Google Scholar 

  16. Iatrou M, Ross SG, Manjeshwar RM, Stearns CW. A fully 3D iterative image reconstruction algorithm incorporating data corrections. Nuclear Science Symposium Conference Record, IEEEVolume 4, Issue 16–22 Oct 2004, pp 2493–7. DOI 10.1109/NSSMIC.2004.1462761.

  17. Seppänen MP, Komar G, Kotoneva E, Teräs M, Minn H. Comparison of 2D and 3D VUE Point reconstruction of PET images with different acquisition times in oncological patients [abstract]. Eur J Nucl Med Mol Imaging 2006;33 Suppl 2:137.

    Google Scholar 

Download references

Acknowledgements

We wish to thank all TPC GE staff involved in this project. We especially thank Helmut Bammer and Eira Kotoneva for helping to perform the phantom studies.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Teräs.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Teräs, M., Tolvanen, T., Johansson, J.J. et al. Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCT. Eur J Nucl Med Mol Imaging 34, 1683–1692 (2007). https://doi.org/10.1007/s00259-007-0493-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-007-0493-3

Keywords

Navigation